Developing a New LIFU Neuromodulation Method to Suppress Tremor

PHASE2Not yet recruitingINTERVENTIONAL
Enrollment

20

Participants

Timeline

Start Date

September 15, 2025

Primary Completion Date

July 15, 2027

Study Completion Date

July 15, 2027

Conditions
Essential Tremor
Interventions
DEVICE

Transcranial focused ultrasound stimulation

Parameters Range of values Fundamental frequency 250 kHz Transducer diameter 15 cm Focal distance 13.5 cm (steerable +/- 5-10 mm from the center) Pulse repetition frequency 100 Hz Duty cycle 10% Duration 2 min for each of the 7 foci Peak spatial intensity (Isppa) 10.0 W/cm2 Peak temporal average intensity (Ispta) 1.2 W/cm2 Mechanical index (MI) 0.48 \* Thermal index (TI) 0.8oC \*\*

DEVICE

Sham stimulation

Same set-up as for real stimulation but with without delivering any ultrasound stimulation