Quantitative monitoring
The TetraGraph TOF monitor will be used to quantitatively assess the TOF-ratio. Quantitative monitoring includes objective measurement of the TOF-ratio, as defined by the ratio between the first and the fourth muscular responses and is quantified by acceleromyography. A TOF twitch response ratio equal or greater to 95% (TOF0.95) will be targeted prior to extubation. If TOF0.95 is not reached within 10 min of administering the initial dose of the respective reversal agent, redosing is recommended.
Qualitative monitoring
Clinicians will assess the TOF-ratio. Qualitative monitoring includes subjective assessment of the count and the amplitude of the muscular responses related to neuromuscular block depth. A TOF twitch response count of 4 similarly strong twitches will be targeted prior to extubation. If 4 similarly strong TOF twitch responses are not reached within 10 min of administering the initial dose of the respective reversal agent, redosing is recommended.