Cleaning Efficacy of Pulsed Vacuum Device on Rigid Endoscopic Instruments

NACompletedINTERVENTIONAL
Enrollment

800

Participants

Timeline

Start Date

July 1, 2024

Primary Completion Date

December 31, 2024

Study Completion Date

December 31, 2024

Conditions
Post-Surgical Contamination of Rigid Endoscopic Instruments
Interventions
DEVICE

Pulsed Vacuum Cleaning and Disinfection Device

After manual pre-treatment, instruments were placed into the pulsed vacuum cleaning and disinfection machine (Shandong Xinhua Medical, Model PC-150L). The automated program included pulsed vacuum cleaning, rinsing, thermal disinfection (93°C for 2.5 minutes), and vacuum drying.

DEVICE

Vacuum Ultrasonic Cleaner

After manual pre-treatment, instruments were cleaned in a vacuum ultrasonic cleaner (Shandong Xinhua Medical, Model super-6000) with a multi-enzyme solution for a minimum of 5 minutes. This was followed by a separate process in a spray cleaning and disinfection machine for rinsing and drying.