264
Participants
Start Date
March 1, 2024
Primary Completion Date
December 31, 2024
Study Completion Date
June 30, 2025
Neuromonitoring of the recurrent laryngeal nerves
The NIM Vital system (Medtronic, Jacksonville, US) operates with surface electrodes integrated with an endotracheal NIM TriVantage tube 7.0-8.0 in diameter, which is inserted by an anesthetist between the vocal folds under direct vision during intubation. The standardized technique of IONM RLNs wll be used, including initial vagal response evaluation at the beginning and final vagal response evaluation at the end of surgery (IONM = L1+V1+R1+R2+V2+L2) according to the recommendations formulated by the International Intraoperative Neural Monitoring Study Group. The nerves will be stimulated using a monopolar electrode and the interrupted stimulation technique at 1mA, 100ms impulse duration and 4Hz frequency. In case of the bifurcated RLN nerves, the assessment includes post-stimulation response of each nerve branch. Adduction of the vocal folds is detected by the endotracheal tube electromyography and abduction by finger palpation of muscle contraction in the posterior cricoarytenoid.
RECRUITING
Department of Endocrine Surgery, Jagiellonian University College of Medicine, Krakow
Marcin Barczynski
OTHER