Saline-induced Distal to Aortic Coronary Pressure Ratio vs. Resting and Hyperemic Indices of Coronary Artery Stenosis Severity

NACompletedINTERVENTIONAL
Enrollment

90

Participants

Timeline

Start Date

January 31, 2020

Primary Completion Date

January 1, 2024

Study Completion Date

January 1, 2024

Conditions
Coronary Stenosis
Interventions
PROCEDURE

Saline Pd/Pa

"Use of a ≥6F guiding catheter and the Verrata pressure guidewire. Administration of intracoronary nitroglycerin.~Verrata normalisation and placement of the Verrata sensor ≥3 vessel-diameter beyond the stenosis.~Whole cycle resting Pd/Pa and resting iFR measurements.~Measurement of contrast FFR: record 5 baseline heart beats, injection of contrast media, 10 ml in the left coronary system, 8 ml in the right coronary system, measure minimal Pd/Pa under stable hyperaemia.~Adenosine FFR: measurement: record 5 baseline heart beats, injection of intracoronary adenosine, 200 mcg in the left coronary system, 100 mcg in the right coronary system, measure minimal Pd/Pa under stable hyperaemia.~Saline-induced Pd/Pa measurement: record 5 baseline heart beats, injection of saline using the ACIST system during 5 heart beats, measure minimal Pd/Pa under stable hyperaemia.~Perform pressure guidewire pullback and check for drift: drift ±0.02 is accepted, if ≥0.03, repeat measurement."

Trial Locations (1)

Unknown

Geneva University Hospitals, Geneva

All Listed Sponsors
lead

University Hospital, Geneva