Accuracy of Non-Invasive Blood Pressure Measurements at the Arm, Calf , and Finger

NACompletedINTERVENTIONAL
Enrollment

133

Participants

Timeline

Start Date

February 22, 2020

Primary Completion Date

June 2, 2021

Study Completion Date

June 2, 2021

Conditions
Hypotension and ShockHypertensionIntensive Care Unit
Interventions
DEVICE

Combined non-invasive intermittent (oscillometric arm cuff), non-invasive continuous (finger cuff and ClearSight device), and invasive continuous (intra-arterial catheter) BP measurements

Continuous (finger cuff) noninvasive and invasive BP measurements, over a 30-min period. Pairs of noninvasive/invasive BP recorded each minute). In the same time, 3 pairs of oscillometric NIBP (cuff placed around a calf), and invasive BP measurements, at 30-sec intervals. The cuff is then placed around an arm and triplets of BP (arm cuff, finger cuff, intra-arterial catheter) will be taken over a 20-min period, at 3-min intervals and every time the invasive and/or the continuous finger cuff measurements cross the threshold of 65 mmHg for the mean BP. If clinically indicated, a cardiovascular intervention that can be either vascular volume expansion or initiation of vasoactive drug therapy, can be administered during this period. Finally, the cuff placed around the arm will be once again placed around the calf to take 3 pairs of oscillometric NIBP measurement with cuff placed around a calf, and simultaneous invasive BP measurement, at 30-second intervals.

Trial Locations (3)

37044

CHRU de Tours, Tours

44093

CHU de NANTES, Nantes

Unknown

CHR Orléans, Orléans

All Listed Sponsors
lead

Centre Hospitalier Régional d'Orléans

OTHER