External Ventricular Drain Placement Stealth Study

NARecruitingINTERVENTIONAL
Enrollment

100

Participants

Timeline

Start Date

August 15, 2018

Primary Completion Date

December 31, 2026

Study Completion Date

December 31, 2026

Conditions
External Ventricular DrainIntracranial Pressure
Interventions
OTHER

Axium Steatlh Image Guidance

The Axiem Stealth imaging guidance system uses a probe (AxiEM Registration Probe) and a side mount emitter (AxiEM Side Mount Emitter) to register the patient's head in relation to the probe. A trajectory (entry point and target) is then planned. This registration process takes an estimated 5-15 minutes to perform. A separate stylet (StealthStation EM Stylet) is designed specifically for the catheter and slides down the center of the catheter. A burr-hole is created at the predefined precise location and the catheter is passed to the target under electromagnetic guidance.

OTHER

EVD placement

A sterile field is prepared about 10-12 centimeters posterior to the eye and about 2-3 cm lateral to the midline of the skull. An incision is made at this point and a burr hole is created using a hand drill. The catheter is advanced in a trajectory perpendicular to the skull, which results in a direction that is towards the ipsilateral medial canthus and ipsilateral tragus which are other landmarks that can be used.