2
Participants
Start Date
August 21, 2018
Primary Completion Date
August 29, 2022
Study Completion Date
August 29, 2022
Ultrahigh resolution Spectral Domain OCT
A spectrometer based ultrahigh resolution Spectral Domain OCT (SDOCT) system operating at 800 nm for the anterior chamber will be employed in the present study. The spectrum of the Ti:Sapphire laser light source is centered at 800 nm. With a full width at half maximum bandwidth of 170 nm, the axial resolution is 1.3 μm in the cornea. The transverse resolution of the employed OCT system is 21 μm at the front surface of the cornea. For measurement, patients will place their head in a modified slit lamp head rest. During the measurement period, patients will be asked to look straight forward onto an internal fixation target and to avoid blinking. Different scattering patterns, e.g. raster, circular and spiral scans will be employed.
Vienna Institute for Research in Ocular Surgery (VIROS), Vienna
Department of Clinical Pharmacology, Medical University of Vienna, Vienna
Medical University of Vienna
OTHER