Heart Failure Study of Multi-site Pacing Effects on Ventriculoarterial Coupling

CompletedOBSERVATIONAL
Enrollment

80

Participants

Timeline

Start Date

September 18, 2017

Primary Completion Date

September 1, 2020

Study Completion Date

September 1, 2020

Conditions
Heart Failure
Interventions
DEVICE

Activation of multi-site pacing capability on CRT devices

"Instead of administering a single LV pulse at the most (electrically) delayed segment of the ventricle, multi-site pacing allows for a more detailed sculpting of the LV activation sequence. Based on the MPP-IDE study results, activation of the antero-lateral wall, or at least its most delayed segments, closely followed by a pulse to the apex and then by a right ventricular one will yield favorable results in terms of hemodynamics and clinical parameters.~Programming features:~* Interpolar distance for the first left ventricular pulse \>30mm (i.e. non-sequential poles used)~* Nearly simultaneous (Δt=5msec) second left ventricular pulse and~* Threshold of ≤[email protected]"

Trial Locations (1)

11527

First Department of Cardiology, Hippokration General Hospital, Athens

All Listed Sponsors
lead

National and Kapodistrian University of Athens

OTHER

NCT03189368 - Heart Failure Study of Multi-site Pacing Effects on Ventriculoarterial Coupling | Biotech Hunter | Biotech Hunter