Using Radiation-free Ultrasound for Screening Scoliosis Among School Children in Hong Kong to Reduce Unnecessary X-ray Exposure

NACompletedINTERVENTIONAL
Enrollment

442

Participants

Timeline

Start Date

September 18, 2017

Primary Completion Date

November 30, 2018

Study Completion Date

May 31, 2019

Conditions
Scoliosis
Interventions
DEVICE

Ultrasound

"Scolioscan, the ultrasound system reported to be reliable and valid for spinal deformity assessment, will be used. The system composes of an ultrasound scanner with a linear probe of 100 mm in width and frequency range of 4-10MHz, a frame structure and a spatial sensor which is attached to the ultrasound probe for spatial data capture. Daily calibration will be performed using a phantom to assure accuracy of spine image formation and subsequent angle measurement.~Subjects will stand on the Scolioscan platform with a standardized posture kept stable with pegs throughout the scanning process.. After adjusting the ultrasound scanner setting, the probe will be steered from L5 to T1 spinous process for scanning.~SPA will be used to determine the ultrasound-based Referral Status through predicting whether the Cobb angle is beyond the referral threshold of ≥20° or not. There are two values for the Referral Status: either for specialist referral or not for specialist referral."

Trial Locations (1)

Unknown

Department of Orthopaedics and Traumatology, Prince of Wales Hospital, Hong Kong