Relationship Between Optimal Laryngeal Mask Airway Cuff Volume and Physical Examination of Head and Neck

NACompletedINTERVENTIONAL
Enrollment

170

Participants

Timeline

Start Date

August 31, 2015

Primary Completion Date

August 31, 2015

Study Completion Date

December 31, 2015

Conditions
Anesthesia
Interventions
DEVICE

Classic laryngeal mask airway

"1. measurement of head and neck~ * Ht,Wt,BMI, neck circumference above a thyroid cartilage, mandible circumference( from the lower part of earlobe-mandible angle-mentum-mandible angle to the lower part of earlobe), neck length (mastoid tip\~sternal notch), modified mallampati classification under sitting position, thyromental distance.~2. laryngeal mask airway insertion and measuring cuff pressure by pressure transducer~ * \# 4.0 for female, #5.0 for male~ * initial cuff volume(air) : maximum volume recommended by the manufacturer~ * check pressure : if it exceed 60cmH20, we will remove air until 60cmH20~ * remove air until gas leak occured around the cuff under airway pressure 15cmH20, fresh gas flow 6liters/min. Check the volume and pressure just before leakage occured."

Trial Locations (1)

136705

Korea university anam hospital, Seoul

All Listed Sponsors
lead

Korea University Anam Hospital

OTHER