5
Participants
Start Date
November 1, 2013
Primary Completion Date
July 31, 2026
Study Completion Date
January 31, 2027
Neural Prosthetic System 2 (NPS2)
The NPS2 comprises 3 NeuroPort Arrays (SIROF). The tip of the electrodes are sputtered iridium oxide film (SIROF). Each array is comprised of 100 1.5 mm microelectrodes organized on a 4mm x 4mm silicon base that is 0.25 mm thick. Each microelectrode is insulated with Parylene-C polymer and is electrically isolated from neighboring electrodes by non-conducting glass. Of the 100 electrodes, 96 are wire bonded using 25m gold alloy insulated wires sealed with a silicone elastomer. The wire bundle is potted to a printed circuit board with epoxy, the circuit board is inserted into the Patient Pedestal (percutaneous connector), and then the Patient Pedestal is filled with silicone elastomer. Two fine platinum reference wires are also attached to the Patient Pedestal. The Patient Pedestal is 19 mm wide at the skin interface.
RECRUITING
University of Colorado Anschutz Medical Campus, Aurora
RECRUITING
University of Southern California, Los Angeles
RECRUITING
Rancho Los Amigos National Rehabilitation Center, Downey
RECRUITING
Richard Andersen, Pasadena
University of Southern California
OTHER
Rancho Los Amigos National Rehabilitation Center
OTHER
University of Colorado, Denver
OTHER
Richard A. Andersen, PhD
OTHER