Conventional Pressure Controlled SIMV
Initial ventilatory settings per clinical physician's discretion-tcCO2 \& Saturation (SAT) monitoring. Adjust vent setting approximately every 15min with a goal of: SAT 88-95%, tcCO2 45-55, pH greater than or equal to 7.25. Once goals reached, ventilatory strategy will crossed over to other intervention.
Mid-frequency Ventilation
Ventilatory parameters include: max possible rate between 61-150 breaths/min; Shortest inspiration time (Ti) and expiration time (Te), but completed inspiration and expiration on pulmonary graphics. Stepwise weaning of PIP by 1 cm H20 to maintain tcCO2 at baseline. If measured PEEP - delivered PEEP generated is \>/= to 1cm H2O, PEEP is reduced to baseline value. Adjustments will be made to PIP, then rate for pH and CO2 changes; FiO2, then PIP for FiO2 changes. Target goal blood gas parameters (same as baseline) Saturations 88-95%; tcCO2 45-55 mmHG; pH \>/= 7.25. Maximum period of 2 hours. When goals met, or 2 hours in period, reverted back to SIMV at baseline settings