Inhaled Dose Analysis Using a Breath Actuated Nebulizer in Healthy Subjects

NACompletedINTERVENTIONAL
Enrollment

25

Participants

Timeline

Start Date

March 13, 2024

Primary Completion Date

April 29, 2024

Study Completion Date

May 1, 2024

Conditions
Healthy VolunteersAerosol TherapyInhaled Drug DosePharmacogenomic Drug Interaction
Interventions
DEVICE

Breath-actuated vibrating mesh nebulizer

The intervention utilizes a breath-actuated vibrating mesh nebulizer system consisting of a controller module with microphone for respiratory phase detection and algorithm for inspiration triggering. This interfaces with the vibrating mesh nebulizer module which incorporates a micro-pump chip using piezoelectric effects to eject fluid through a mesh aperture plate holes. By detecting the onset of inspiration based on characteristic acoustic patterns using machine learning models, the controller module sends signals activating the piezoelectric vibration mechanism to generate aerosol only during the inspiratory phase through precision timing control, shutting off mist during expiration. The core module is attached to a standard commercial vibrating mesh nebulizer using the standard adult reusable mouthpiece interface.

DEVICE

Continuous vibrating mesh nebulizer

When the continuous vibrating mesh nebulizer powered on, the micro-pump chip provides sustained vibrations onto the aperture plate, forcing fluid through micron-scale pores under pressure to continuously produce aerosol throughout inspiration and expiration phase.

Trial Locations (1)

33305

Linkou Chang Gung Memorial Hospital, Taoyuan District

All Listed Sponsors
collaborator

National Tsing Hua University

OTHER

collaborator

Chang Gung Memorial Hospital

OTHER

collaborator

National Science and Technology Council

FED

lead

Chang Gung University

OTHER

NCT06157515 - Inhaled Dose Analysis Using a Breath Actuated Nebulizer in Healthy Subjects | Biotech Hunter | Biotech Hunter