25
Participants
Start Date
March 13, 2024
Primary Completion Date
April 29, 2024
Study Completion Date
May 1, 2024
Breath-actuated vibrating mesh nebulizer
The intervention utilizes a breath-actuated vibrating mesh nebulizer system consisting of a controller module with microphone for respiratory phase detection and algorithm for inspiration triggering. This interfaces with the vibrating mesh nebulizer module which incorporates a micro-pump chip using piezoelectric effects to eject fluid through a mesh aperture plate holes. By detecting the onset of inspiration based on characteristic acoustic patterns using machine learning models, the controller module sends signals activating the piezoelectric vibration mechanism to generate aerosol only during the inspiratory phase through precision timing control, shutting off mist during expiration. The core module is attached to a standard commercial vibrating mesh nebulizer using the standard adult reusable mouthpiece interface.
Continuous vibrating mesh nebulizer
When the continuous vibrating mesh nebulizer powered on, the micro-pump chip provides sustained vibrations onto the aperture plate, forcing fluid through micron-scale pores under pressure to continuously produce aerosol throughout inspiration and expiration phase.
Linkou Chang Gung Memorial Hospital, Taoyuan District
National Tsing Hua University
OTHER
Chang Gung Memorial Hospital
OTHER
National Science and Technology Council
FED
Chang Gung University
OTHER