176
Participants
Start Date
June 27, 2022
Primary Completion Date
June 30, 2025
Study Completion Date
August 31, 2025
Cardiac resynchronization therapy
"Standard BiV-CRT The CRT-P/CRT-D device and leads should be implanted according to the physician's manual provided with the devices. Only locally approved Medtronic CRT-P/CRT-D generators will be used during the study. The LV lead should be implanted at a lateral or posterolateral CS branch in a basal or mid ventricular position confirmed by ortogonal fluoroscopic views (RAO and LAO 40º). Acceptable LV pacing threshold should be \<3V@0,5ms and phrenic nerve stimulation (PNS) margin should be \>1V with respect to the pacing threshold.~LBBAP-CRT The CRT-P/CRT-D device and leads should be implanted according to the physician's standard practice. Only locally approved Medtronic CRT-P/CRT-D generators will be used during the study. During the implant procedure lead impedances, pacing and sensing parameters will be measured using a pacing system analyzer (PSA). For LBBAP, the Medtronic 3830 lead will be used. Acceptable LBBAP threshold should be \<2,5V@0,5ms"
Hospital Universitario Virgen de las Nieves, Granada
Hospital Universitario San Cecilio, Granada
Hospital Universitario Juan Ramón Jiménez, Huelva
Hospital 12 Octubre, Madrid
Hospital Universitario Puerta De Hierro, Majadahonda
Hospital Universitari i Politècnic La Fe, Valencia
Hospital Clínico Universitario Lozano Blesa, Zaragoza
Hospital General Universitario de Alicante Doctor Balmis, Alicante
Hospital Universitario Virgen de la Arrixaca, Murcia
Hospital Universitario de Bellvitge, L'Hospitalet de Llobregat