101
Participants
Start Date
August 31, 2012
Primary Completion Date
November 30, 2013
Study Completion Date
November 30, 2013
NIPPV
Ventilation with nasal intermittent positive pressure - Newborns randomized to this group underwent NIPPV, the respirators in TIME mode, set with the following settings : 5-6 centimeters of water (cmH2O) pressure positive end expiratory (PEEP); Flow rate of 6-8 liters per minute (L / min); PIP(positive inspiratory pressure) \< 15 centimeters of water (cm H2O) ; respiratory frequency backup (FR) \< 16 bpm; inspiratory time (IT) - 0.30 to 0.33 seconds (adjustable time constant according to the weight and baseline disease)
Bubble NCPAP
Bubble NCPAP - Newborns randomized to this group bubble NCPAP, adjusted with the following settings: pressure positive end expiratory (PEEP) 5-6 centimeters of water (cm H2O) and Flow 6-8 liters per minute (L / min), provided through oxygen sources and compressed air (15 liters per minute - L / min), which passes through a heated humidifier through a breathing circuit that goes to the newborn, and from him, for a plastic container (250 milliliters (ml) capacity and 14 centimeters) containing sterile distilled water, which is the bubble.
Ventilator NCPAP
Ventilator NCPAP - Newborns randomized to this group ventilator NCPAP, of respirators in TIME mode set with the following settings: pressure positive end expiratory (PEEP) 5-6 centimeters of water (cm H2O) and Flow 6-8 liters per minute (L / min).
Federal University of Minas Gerais
OTHER