Standard Defibrillation Testing
Once the ICD is implanted, Standard Defibrillation Threshold (DFT) testing will begin. A low energy shock will be delivered to the heart at the peak of the T wave to induce ventricular fibrillation (VF). Once VF is induced, a rescue shock of 25J will be delivered to restore sinus rhythm. This process will be repeated once more after 5 minutes.
Upper Limit of Vulnerability Testing
Once the ICD is implanted, Upper Limit of Vulnerability (ULV) testing will commence. An 18 J shock will be delivered will at different points from the t-wave. If VF is not induced with any of these shocks, then the ULV will be considered to be 18J. If VF is induced, then a 25 J rescue shock will be delivered.