20
Participants
Start Date
September 30, 2011
Primary Completion Date
December 31, 2014
Study Completion Date
December 31, 2014
Blood sample for bupivicaine pharmacokinetics
"Bupivacaine levels will be measured with an assay developed on the 4000 QTrap Tandem Mass Spectrometer (Applied Biosystems/ MDS Sciex).~Data will be collected and analyzed using the Agilent ChemStation software (Rev A.10.02). The analytical column is an Agilent Zorbax Eclipse XDB-C8 (4.6 x 150mm, 5µm).Quantification of bupivacaine is based on analyzing a six point calibration curve (0 to 4.0 mg/L). An internal standard (Prilocaine 221.2/86.1) is included with each calibrator, quality control or patient sample. Instrumentation and Chromatographic conditions: The HPLC system consists of a Specta-System P1000 Isocratic Pump and an AS3000 Autosampler (Thermo Electron Corporation, Waltham MA)."
The Hospital for Sick Children, Toronto
The Hospital for Sick Children
OTHER