vv-ECCO2R (Novalung GmbH, Germany)
Treatment with the extracorporeal lung assist (ECLA) The ECLA is a pump driven (centrifugal pump) venovenous circuit, which removes carbon dioxide from the patients blood by means of a membrane through which the patients blood runs on the one side of the membrane and sweep gas on the other side removing the patient´s carbon dioxide. Blood flow range from 0.5 to 4.5 l/min and sweep gas flow between 1 and 10 l/min. At blood flows of 2 l/min and higher the device also oxygenates the patients blood. The diameter and length of cannulas and the sites of venous insertions are left to the decision of the treating physician. Cannulas are inserted in seldinger technique under sterile conditions. Function and patency of the extracorporeal circuit requires mild therapeutic anticoagulation.