ENHANCED Device Programming to Reduce Therapies and Improve Quality of Life in Implantable Cardioverter Defibrillator (ICD) Patients

NAUnknownINTERVENTIONAL
Enrollment

60

Participants

Timeline

Start Date

April 30, 2013

Primary Completion Date

April 30, 2016

Study Completion Date

April 30, 2016

Conditions
Cardiac Arrhythmias
Interventions
DEVICE

Enhanced ICD programming

"* VT monitor: \> 166/min~* fVT: \> 182/min; via VF 60/80 intervals (number of intervals to start ATP after approximately 20 s); 3 x ATP (8 stimuli, 88%, scan 20 ms); shock 1-5: 35 J; redetection 30/40 intervals.~* VF: \> 250/min; via VF 60/80 intervals (number of intervals to start therapy after approximately 15 s); 1 x ATP (8 stimuli, 88%) during charging; all shocks: 35 J; redetection 30/40 intervals~* SVT/VT discrimination is turned on, high rate time out is OFF~* SVT/VT discrimination single chamber: stability, wavelet; SVT upper rate limit: 222/min~* SVT/VT discrimination dual/triple chamber: P/R logic, wavelet; SVT upper rate limit: 222/min~* T wave-oversensing and lead noise discrimination is turned on in all devices."