56
Participants
Start Date
November 30, 2011
Primary Completion Date
December 31, 2011
Study Completion Date
February 29, 2012
Prescription of post-dialytic weight based on BCM device
"Prescription of post-dialytic weight target: On a weekly base, the patients are pre-dialytically measured by BCM (body composition monitor).~The time average fluid overload (TAFO) determines the target range. TAFO is defined as the mean value of pre- and post-dialytic overhydration (TAFO = (OHpre+OHpost)/2). The target range is set asymmetrically around a TAFO of 0.8, from -1.0 L to +0.5, i.e. the TAFO target range will be from -0.2 L to 1.3 L.~Prescription steps will be calculated weekly for all study patients. TAFO \< -0.2 L: Increase post-dialytic weight by Cw x 0.5 kg/week, TAFO between -0.2 to 1.3 L: No action, TAFO between 1.3 to 2.8 L Decrease post-dialytic weight by Cw x 0.5 kg/week, TAFO \> 2.8 L: Decrease post-dialytic weight by Cw x 1.0 kg/week."
Dialysis Units, Hospital Clínic, Barcelona
Fresenius Medical Care Europe
UNKNOWN
Francisco Maduell
OTHER