Clinical Investigation of Wireless Transmission of Invasive Blood Pressure Signal

CompletedOBSERVATIONAL
Enrollment

12

Participants

Timeline

Start Date

June 30, 2011

Primary Completion Date

June 30, 2011

Study Completion Date

June 30, 2011

Conditions
Multiple TraumaShockAtrial FibrillationPerioperative/Postoperative Complications
Interventions
DEVICE

Wireless invasive blood pressure monitoring (Hybymed® Wireless IBPM System)

"The readings of invasive arterial pressure were obtained from 2 identical pressure transducers, connected to a common arterial line via a Y connector. Data was collected from one transducer via a direct cable connection and from the second transducer via wireless transmission of the signal to the patient's bedside monitor using the HMW 10 Wireless System.~A comparison of data received by patient's bedside monitor from the two transducers was done (standard connection vs. wireless connection), as well as a comparison of data collected by the transmitter and data received by the receiver of the HMW 10 Wireless System."