Clinical Evaluation of the Accuracy of the FreeStyle Navigator Continuous Glucose Monitoring System

CompletedOBSERVATIONAL
Enrollment

58

Participants

Timeline

Start Date

November 30, 2005

Primary Completion Date

December 31, 2005

Study Completion Date

December 31, 2005

Conditions
Diabetes Mellitus
Interventions
DEVICE

FreeStyle Navigator Continuous Glucose Monitoring System

Subjects wore 2 Navigator sensors for approximately 122 hours. Subjects and clinical staff were masked to continuous glucose data. During the time the subjects wore the continuous sensor, they were instructed to continue with their current diabetes management plan. Subjects stayed at the clinic for approximately 50 hours during the study period. While in the clinic, venous blood samples were collected every 15 minutes to determine whole blood reference glucose concentrations using the Yellow Springs Instrument. Subjects were also asked to obtain up to 50 FreeStyle finger blood glucose measurements using the system in the Navigator receiver for comparative analysis and system calibration. This was not a treatment study.

Trial Locations (3)

78229

Diabetes & Glandular Disease Clinic, San Antonio

94598

Diablo Clinical Research, Walnut Creek

98055

Ranier Clinical Research, Renton

Sponsors
All Listed Sponsors
lead

Abbott Diabetes Care

INDUSTRY

NCT00920881 - Clinical Evaluation of the Accuracy of the FreeStyle Navigator Continuous Glucose Monitoring System | Biotech Hunter | Biotech Hunter